Filter sand mill, which is more favored by high viscosity materials, gets rid of the traditional blockage problem and is equipped with Boyi TurboMax independently developed turbine structure.
Provide excellent operational stability, higher equipment flow, improve production efficiency and reduce costs
Resistance | ||
RANGE | ACCURACY | RESOLUTION |
0.01Ω~0.999Ω | ±(1.5%+0.01Ω) | 0.001Ω |
1Ω~9.99Ω | ±(1.5%+0.1Ω) | 0.01Ω |
10Ω~99.9Ω | ±(2.0%+0.3Ω) | 0.1Ω |
100Ω~199.9Ω | ±(3.0%+1Ω) | 1Ω |
200Ω~400Ω | ±(6.0%+5Ω) | 5Ω |
400Ω~500Ω | ±(10%+10Ω) | 10Ω |
500Ω~1200Ω | Approx.20% | 20Ω |
Current | ||
RANGE | ACCURACY | RESOLUTION |
100mA | ±(2.5%+1mA) | 0.1mA |
300mA | ±(2.5%+2mA) | 0.3mA |
1A | ±(2.5%+0.003A) | 0.001A |
3A | ±(2.5%+0.01A) | 0.003A |
10A | ±(2.5%+0.03A) | 0.01A |
20A | ±(2.5%+0.05A) | 0.03A |
30A | ±(2.5%+0.5A) | 0.03A |
40A | ±(10%+0.5A) | 0.03A |
Test conditions | ||
Temperature | 23℃±3℃ | |
Humidity | 50%RH±10% | |
Power supply of battery | >7V | |
External magnetic field | <40A/m | |
External electric field | <1V/m | |
Current test frequency | 45Hz∽65Hz | |
Technical Specification | ||
Single measurement time | 1s | |
Test voltage | 3700V | |
Mean consumption | 50mA | |
Electric shock resistance | IEC1010-1 | |
Range | automatic | |
Power shortage | display symbol | |
Overload limit | normal current 40A RMS | |
Diameter of jaw | Φ32mm or 45mm×32mm | |
Working temperature | -10℃∽50℃(14℉ to 122℉) | |
Storage temperature | -20℃∽60℃(-4℉ to 140℉) | |
Display type | 4 digits, 9999 counting, liquid crystal display | |
Electric clearance | 6.5mm (meet IEC1010 double insulation CATⅡ 600V) | |
GENERAL | ||
Power supply | 9V 6F 22 | |
Outline size | 270mm×96mm×55mm | |
Net weight | 1002g(not include batteries) | |
Gross weight | 2574g | |
Safety Rating | EN61010-1,EN61557-3,EN61326-1,CAT.Ⅱ600v |
The 2301 is a big breakthrough of traditional grounding resistance measurement technology, and is used for measuring the grounding resistance of the system with circuits, for example, the grounding resistance of power transmission lines or the grounding resistance of communication lines. Auxiliary grounding rods are not needed to use when measuring, several parallel connected grounding system can be applied, and the equipment to be tested are not interrupted. The grounding resistamce can be safety and quickly measured by champing the grounding wires or the grounding rods only with the jaws.
Resistance detector ring (1Ω) | One piece |
Resistance detector ring (5Ω) | One set |
Resistance detector ring (10Ω) | One piece |
9V battery | Four pieces |
Metal carry case | One piece |
Gift box | One piece |
We aim to produce intelligent instruments, provide high quality services, and become an innovative manufacturer.
Our R&D team have over 20 years R&D experience in instrument industry .
Our mission is to get our work done well to become customer’s reliable business friend. S.E.Q( Service , Economic pricing , Quality ) is our concern , we know what you need ,and we know what we have .